Mark Event
If an event is marked A, B, C or D, the mark ( A, B, C or D ) will be displayed aside the
mark time on the trend table.
9.4 NIBP Recall
Select NIBP RECALL >> in SYSTEM MENU. The following window appears.
Figure 9-3 NIBP Recall
The NIBP RECALL window shows the non-invasive systolic pressure (NS), non-invasive
mean pressure (NM), non-invasive diastolic pressure (ND), puse rate (PR) and the
measurement time (TIME). The optional pressure units (UNIT) are mmHg and kPa. NUM
indicates the current measurement times. PR is derived from NIBP measurements. A
maximum of ten measurements can be displayed on the screen once. If there are more, you
may use the UP-DOWN option to review the
data of a later or earlier time. If the measurement times surpass 800, only the latest 800
measurements are to be displayed. The REC option allows you to print out all measurement
data of NIBP RECALL.