4.4.6 Analog Output
Select ANALOG >> in SYSTEM SETUP menu. The following menu appears.
Figure 4-12 Analog Output
You can perform the following settings in the menu above:
Analog output
ANALOG OUT
Options: ON and OFF.
When ON is selected, analog signals can be output from
the auxiliary output port on the rear panel of the monitor.
Options: ECG1, ECG2, IBP1 and IBP2;
ANALOG WAVE
In the MULTI-LEADS DISPLAY mode, the ECG3,
ECG4, ECG5 and ECG6 options are also available.
NOTE
If DEFIB. SYN or NURSE CALL is selected from the AUX OUTPTU options in
the USER MAINTAIN menu, the ANALOG>> option in the SYSTEM SETUP
menu will be inactive, and the monitor will be unable to output analog signals. For
details, please refer to 4.7 Maintenance.