15.4.3 IBP Pressure Calibration
You can select IBP PRESSURE CALIBRATE in IBP (1,2) SELECT menu to open the
following menu.
Figure 15-11 IBP Pressure Calibration Menu
Calibration Procedure
1.
Disconnect the pressure transducer from the patient. Connect the 3-way stopcock, the
sphygmomanometer and the inflation orb, using a T-shape connector, as shown below.
Figure 15-12 IBP Pressure Calibration
3-way stopcock
Pressure
transduce
Pressure transducer
interface cable
Monitor Hydrargyrum
pressure meter
T-shape connector
15-10